114
90000 A-Series Service Guide
5
Replacing Assemblies
Removing and replacing the power cord
1
Disconnect the power cable and remove the cover and top plate.
2
Unplug the power cable from the +12 V power supply.
3
Unscrew the power cord connector from the rear panel (Torx T10 screws).
4
Remove the ground screw (Torx T10) on right side of hard drive.
5
Pull the power cord out of the instrument.
6
To replace the power cord, reverse this procedure.