44
90000 A-Series Service Guide
3
Testing Performance
DC Gain Accuracy Test
Vertical Scale
V
Aux Out
Setting
V
DMM+
V
Scope+
V
DMM-
V
Scope-
Calc. DC
Gain Error
DC Gain Error
Test Limits
Channel 1
5 mV/div
±15 mV
±4.5%
10 mV/div
±30 mV
±2%
20 mV/div
±60 mV
±2%
50 mV/div
±150 mV
±2%
100 mV/div
±300 mV
±2%
200 mV/div
±600 mV
±2%
500 mV/div
±1.5 V
±2%
1 V/div
±2.4 V
±2%
Channel 2
5 mV/div
±15 mV
±4.5%
10 mV/div
±30 mV
±2%
20 mV/div
±60 mV
±2%
50 mV/div
±150 mV
±2%
100 mV/div
±300 mV
±2%
200 mV/div
±600 mV
±2%
500 mV/div
±1.5 V
±2%
1 V/div
±2.4 V
±2%
Channel 3
5 mV/div
±15 mV
±4.5%
10 mV/div
±30 mV
±2%
20 mV/div
±60 mV
±2%
50 mV/div
±150 mV
±2%
100 mV/div
±300 mV
±2%
200 mV/div
±600 mV
±2%
500 mV/div
±1.5 V
±2%
1 V/div
±2.4 V
±2%
Channel 4
5 mV/div
±15 mV
±4.5%
10 mV/div
±30 mV
±2%
20 mV/div
±60 mV
±2%
50 mV/div
±150 mV
±2%
100 mV/div
±300 mV
±2%
200 mV/div
±600 mV
±2%
500 mV/div
±1.5 V
±2%
1 V/div
±2.4 V
±2%
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......