82
90000 A-Series Service Guide
5
Replacing Assemblies
Removing and replacing the cover, top plate, and bottom plate
1
Disconnect the power cable, all probes and BNC input cables, the accessory
pouch if it is attached, and any other cables including those for a mouse,
keyboard, printer, USB, and LAN.
2
Remove the two Torx T20 screws securing each side handle. Standard models
have one handle; performance-enhanced models have two handles.
3
Remove the 6 Torx T20 screws that secure the rear feet (three in each foot).