16
90000 A-Series Service Guide
2
Calibration
Procedure
1
Let the oscilloscope warm up before running the self calibration.
The self calibration should be done only after the oscilloscope has run for 30
minutes at ambient temperature. Calibration of an oscilloscope that has not
warmed up may result in inaccurate calibration.
2
Click
Utilities > Calibration...
.
3
Uncheck the
Cal Memory Protect
box. You cannot run a self calibration if this box
is checked.
4
Click
Start Full User Cal
, then follow the instructions on the screen. The routine
will prompt you to follow these steps:
a
Disconnect everything from all inputs and Aux Out.
b
Select the level of calibration you want to perform.
•
Standard Calibration—Time scale calibration will not be performed. Time
scale calibration factors from the previous time scale calibration will be used
and the 10 MHz reference signal will not be required. The remaining
calibration procedure will continue.
•
Standard Calibration and Time Scale Calibration—Time scale calibration will
be performed. This option requires you to connect a 10 MHz reference signal
to channel 1 that meets the following specifications. Failure to use a
reference signal that meets these specifications will result in an inaccurate
calibration.
Frequency: 10 MHz ±0.4 ppm = 10 MHz ±4 Hz
Amplitude: 0.2 Vpeak-to-peak to 5.0 Vpeak-to-peak
Wave shape: Sine or Square
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......