112
90000 A-Series Service Guide
5
Replacing Assemblies
Removing and replacing the power supply
1
Disconnect the power cable and remove the top cover.
2
Disconnect the AC power cable from the bulk +12 V power supply.
The picture below shows where the AC power cable connects to the bulk 12V
power supply.
3
Follow the instructions on the label attached to the top of the power supply to
remove it.
WARNING
SHOCK HAZARD! If the power supply is defective it could have a dangerous charge
on some capacitors. This charge could remain for many days after removing power
from the supply.
WARNING
SHOCK HAZARD! When the power supply is removed, two AC leads are exposed. Be
careful not to come into contact with these leads.