Troubleshooting
4
90000 A-Series Service Guide
51
4
Turn each knob in both directions until you have turned all knobs. When you
turn a knob in one direction, half the rotation arrow under the corresponding
knob symbol should turn green. When you then turn the knob in the other
direction, the entire rotation arrow under the knob symbol should turn green.
5
When you are finished, click
Close
.
6
If any of the knobs or keys do not work, go to “
” later
in this chapter.
To test the front panel LED indicators:
1
Click
Utilities > Self Test...
.
2
Select
LED Test
from the drop-down list box, then click
Start
.
The Front Panel LED Test window appears, showing a symbolic representation
of all front panel LED indicators.
3
Repeatedly press the
[Single]
button on the front panel to step through and
highlight each LED symbol in the test screen. You can also step through the
LEDs by pressing the
<<Prev
or
Next>>
buttons on the display screen. Verify that
the corresponding LEDs on the front panel are the only ones illuminated.
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......