Replacing Assemblies
5
90000 A-Series Service Guide
115
Removing and replacing the USB or GPIB port
Use this procedure to remove and replace the USB or GPIB port. The GPIB is an
option and may not be on your oscilloscope. The procedure is exactly the same for
both of these ports so only one is shown here.
1
Disconnect the power cable and remove the cover and top plate.
2
Remove the screw from top edge of the chassis.
3
Undo the connection to the motherboard by disconnecting the cable.
4
Pull the port out of the instrument.
5
To replace one of these ports, reverse this procedure.