50
90000 A-Series Service Guide
4
Troubleshooting
Run the oscilloscope self tests
Running the oscilloscope self tests performs a series of internal procedures to
verify that the oscilloscope is working properly.
1
Click
Utilities > Self Test...
.
2
Select
Scope SelfTest
from the Available Self Tests drop-down list box.
3
Click
Start
and follow the instructions on the screen.
If any of the self tests fail, go to the “Acquisition/Backplane assembly
troubleshooting” section.
Run the keyboard, LED, and touch screen self tests
To verify correct keyboard operation:
1
Click
Utilities > Self Test...
.
2
Select
Keyboard Test
from the drop-down list box, then click
Start
.
The Front Panel Keyboard Test window appears, showing a symbolic
representation of the keyboard.
3
Press each key on the keyboard until you have pressed all keys. When you press
a key or push a knob, the bar above the corresponding key/knob symbol on the
display should change from red to green.
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......