Replacing Assemblies
5
90000 A-Series Service Guide
87
Removing and replacing the front panel bezel and front panel circuit
assembly
1
Disconnect the power cable and remove the cover, top plate, and bottom plate
as described in the previous section.
2
Disconnect front panel cables.
3
Grasp and pull on all 13 knobs located on the front panel to remove them.
4
Use a screwdriver to disengage the locking tabs located along the edges of the
front panel bezel. For best results, disengage the four side tabs first, then the
top tab, and then pull the bezel off at an angle and flex it until the bottom tab
disengages.