Calibration
2
90000 A-Series Service Guide
17
•
Standard calibration and Default Time Scale—Factory time scale calibration
factors will be used. The 10 MHz reference signal will not be required. The
remaining calibration procedure will continue.
c
Connect the calibration cable from Aux Out to channel 1.
You must use the 54855-61620 cable assembly with two connector saver
adapters for all oscilloscopes except the DSO90254A, which can use a
good-quality BNC cable. Failure to use the appropriate calibration cable will
result in an inaccurate calibration.
d
Connect the calibration cable from Aux Out to each channel input as
requested.
e
Connect the 50
Ω
BNC cable from the Aux Out to the Aux Trig on the front
panel of the oscilloscope.
f
A Passed/Failed indication appears for each calibration section. If any
section fails, click the Enable Details box for information on the failures. Also
check the calibration cables and run the oscilloscope self tests.
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......