28
90000 A-Series Service Guide
3
Testing Performance
DC Gain Accuracy Test
CAUTION
Ensure that the input voltage to the oscilloscope never exceeds
±5 V
.
NOTE
The oscilloscope under test must be warmed up (with the oscilloscope application running) for
at least 30 minutes prior to the start of any performance test.
Specifications
DC Gain Accuracy
±2% of full scale at full resolution channel scale.
±4.5% of full scale at 5 mV/div.
Full scale is defined as 8 vertical divisions. Magnification is used below 5 mV/div. Below 5 mV/div, full
scale is defined as 40 mV. The major scale settings are 5 mV/div, 10 mV/div, 20 mV/div, 50 mV/div,
100 mV/div, 200 mV/div, 500 mV/div, and 1 V/div.
Description
Critical Specifications
Recommended Model/ Part #
Digital Multimeter
DC voltage measurement accuracy better than
±0.1% of reading
Keysight 34401A
Cable Assembly
(2 required)
50
Ω
characteristic impedance, BNC (m)
connectors
Keysight 8120-1840
Adapter
BNC Tee (m)(f)(f)
Keysight 1250-0781
Adapter
BNC (f) to dual banana
Keysight 1251-2277
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......