B-42
BOP-1K 091719
B.168 SYSTem:MDNS:NAME? QUERY
SYST:MDNS:NAME?
[ME MODELS ONLY]
Syntax:
Short Form: SYST:MDNS:NAME? Long
Form:
SYSTem:MDNS:NAME?
Returns: string up to 12 ASCII characters
Description:
Returns MDNS host name.
May be up to 12 ASCII characters.
B.169 SYSTem:MODE COMMAND
SYST:MODE
Syntax:
Short Form: SYST:MODE {BIP | UNIP}
Long Form: SYSTem:MODE {BIPolar | UNIPolar}
Description:
Used to set mode to Bipolar (default) or Unipolar.
Set to BIPolar to allow the unit to meet the 60
millisecond (maximum) specification between setpoint change and measurement response.
NOTE: When set to BIPolar, changing of the voltage setpoint in voltage mode
requires
using
VOLT:MODE command to set voltage and CURR:PROT to set the current limit. Changing of
the current setpoint in current mode
requires
using CURR:MODE command to set current
and VOLT:PROT to set the voltage limit. When set to BIPolar no drivers are available.
When set to UNIPolar the unit still synchronizes setpoint changes with measurements, but the 60 mil-
liseconds (maximum) specification is not attainable. When set to UNIPolar the unit functions with the
drivers described in the manual.
B.170 SYSTem:MODE? QUERY
SYST:MODE?
Syntax:
Short Form: SYST:MODE?
Long Form: SYSTem:MODE?
Return Value: BIP = Bipolar, UNIP = Unipolar
Description:
Returns BIP or UNIP to indicate whether the mode is set to Bipolar or Unipolar, respectively).
B.171 SYSTem:PASSword:CENable COMMAND
SYST:PASS:CEN
Syntax
:
Short Form: SYST:PASS:CEN <val>
Long Form: SYSTem:PASSword:CENable <val>
Description:
Sets the password enable state if the value matches the current Main password.
This command
allows CALibrate commands to operate.
B.172 SYSTem:PASSword:CDISable COMMAND
SYST:PASS:CDIS
Syntax
:
Short Form: SYST:PASS:CDIS
Long Form: SYSTem:PASSword:CDISable
Description:
Clears the password enable state.
B.173 SYSTem:PASSword:NEW COMMAND
SYST:PASS:NEW
Syntax
:
Short Form: SYST:PASS:NEW <old password>,<new password>
Long Form: SYSTem:PASSword:NEW <old password>,<new password>
Description:
Establishes new password.
The old (current) Main password is checked, then replaced by the new
password. Admin 1 and 2 passwords (PAR. 3.2.4.2) are not available. (See example, Figure B-14.)
See PAR. 3.5.2.2 and Figure 3-19 for special programming considerations.
B.174 SYSTem:PASSword:STATe? QUERY
SYST:PASS:STAT?
Syntax
:
Short Form: SYST:PASS:STAT?
Long Form: SYSTem:PASSword:STATe?
Return Value: <int_value> 0 or 1
Description: Returns a 1 if the password state is enabled or a 0 if it is disabled.
B.175 SYSTem:REMote
COMMAND
SYST:REM
Syntax
:
Short Form: SYST:REM {ON | OFF} or {1 | 0}
Long Form: SYSTem:REMote {ON | OFF} or {1 | 0}
Description:
Used during serial (RS 232) communication to set the unit to remote (1 or ON) or local (0 or
OFF) mode.
This command must be issued prior to commands that affect the power supply output
(e.g., VOLT 10;:OUTP ON) to ensure the unit is in remote mode. See PAR. 3.5.6.3 and Figure B-13.
Summary of Contents for BOP-ME 1KW
Page 2: ......
Page 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Page 42: ......
Page 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Page 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Page 76: ......
Page 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Page 164: ......