3-62
BOP HIPWR 091719
3.5.4
IEEE 488 (GPIB) BUS PROTOCOL (MG MODELS ONLY)
Table 3-19 defines the interface capabilities of the BOP 1KW power supply (Talker/Listener) rel-
ative to the IEEE 488 (GPIB) bus (reference document
ANSI/IEEE Std 488: IEEE Standard Digital
Interface for Programmable Instrumentation
) communicating with a Host Computer—Controller
(Talker/Listener). Tables 3-20 and 3-21 define the messages sent by the BOP, or received by the
BOP, via the IEEE 488 bus in IEEE 488 command mode and IEEE 488 data mode, respectively.
These messages are enabled during the “handshake” cycle, with the BOP power supply operat-
ing as either a Talker or a Listener. (See PAR. 3.5.3.2 to change the GPIB address.
CAUTION: To ensure correct operation, do not remove the GPIB cable from the BOP
after a Query is processed. Ensure that the last command is a non-query
such as *CLS before removing the GPIB cable. Otherwise, power to the BOP
and/or the host computer may need to be cycled OFF then ON to restore
proper operation.
TABLE 3-19. IEEE 488 (GPIB) BUS INTERFACE FUNCTIONS
FUNCTION
SUBSET
SYMBOL
COMMENTS
Source Handshake
SH1
Complete Capability (Interface can receive multiline messages)
Acceptor Handshake
AH1
Complete Capability (Interface can receive multiline messages)
Talker
T6
Basic talker, serial poll, unaddress if MLA (My Listen Address) (one-byte address)
Listener
L4
Basic listener, unaddress if MTA (My Talk Address) (one-byte address).
Service Request
SR1
Complete Capability. The interface sets the SRQ line true if there is an enabled ser-
vice request condition.
Remote/Local
RL1
Complete capability. Interface selects either local or remote information. In local
mode the BOP executes front panel commands, but can be set to remote mode via
IEEE 488 bus. When in Remote mode all front panel keys are disabled except
LOCAL function key. LOCAL function key can be disabled using keypad lockout
command (see Appendix B, PAR. B.162) so that only the controller or a power on
condition can restore Local mode.
Parallel Poll
PP0
No Capability
Device Clear
DC1
Complete Capability. BOP accepts DCL (Device Clear) and SDC (Selected Device
Clear).
Device Trigger
DT1
Respond to *TRG and <GET> trigger functions.
Controller
C0
No Capability
TABLE 3-20. IEEE 488 (GPIB) BUS COMMAND MODE MESSAGES
MNEMONIC
MESSAGE
DESCRIPTION
COMMENTS
ATN
Attention
Received
DAC
Data accepted
Received or Sent
DAV
Data Valid
Received or Sent
DCL
Device Clear
Received (see PAR. 3.5.3.2.2)
GTL
Go To Local
Not Supported. Use SYST:REM 0 instead
(see PAR. B.175)
IFC
Interface Clear
Received
Summary of Contents for BOP-ME 1KW
Page 2: ......
Page 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Page 42: ......
Page 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Page 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Page 76: ......
Page 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Page 164: ......