3-70
BOP HIPWR 091719
FIGURE 3-22. WEB INTERFACE HOME PAGE (UNIT DESCRIPTION)
3.5.7.5
LAN CONFIGURATION USING WEB INTERFACE
From the Web Interface Home Page (Figure 3-22) click on CONFIGURE LAN at the left to view
the LAN Configuration page (Figure 3-23). The parameters that can be configured from this
page are Host Name (DHCP), Description, IP Address, Subnet Mask address, Default Gateway
Address, and DNS Server Address. Separate check-boxes allow enabling of DHCP, AUTOIP
and PING. Descriptions of these terms can be viewed by clicking the HELP box at the left. If a
data entry fault condition exists, an alert window indicates the type of fault. Click OK to close the
window; the SUBMIT button is disabled until the fault is corrected.
There are three passwords shown on this page which control access to the web pages only. All
three web passwords described below are case sensitive.
NOTE: In cases where multiple users are expected to have access to the units, it is recom-
mended that password access be used to restrict
control
of the unit to one user (func-
tioning as a bus controller), while allowing other users viewing access. Allowing
multiple users to issue commands to a single unit can cause unpredictable results and
may cause the unit to lock up.
•
The OPERATE INSTRUMENT password restricts the ability to modify the OPERATE
INSTRUMENT page settings. The settings can still be viewed while password protected
•
The CONFIGURE LAN password restricts the ability to change the LAN configuration
parameters shown on the CONFIGURE LAN page (Figure 3-23.)
Summary of Contents for BOP-ME 1KW
Page 2: ......
Page 20: ...xii BOP 1K 091719 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Page 42: ......
Page 64: ...2 22 BOP HIPWR 091719 FIGURE 2 11 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Page 65: ...BOP HIPWR 091719 2 23 FIGURE 2 12 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Page 76: ......
Page 161: ...BOP HIPWR 091719 3 85 FIGURE 3 29 STATUS REPORTING STRUCTURE...
Page 164: ......