Preventative Maintenance
This pump must be periodically re-certified to assure proper functioning and safety. The recommended service interval is at
least once per year. Re-certification may be done at the user’s Biomedical Engineering Department, an outside service, or
by Tyco Healthcare Group LP Factory Service. To arrange for Tyco Healthcare Group LP Factory Service, call 1-800-448-0190.
In Canada, call 1-877-664-TYCO(8926)
If a pump malfunctions, please contact your THGLP Representative or call Customer Service for instruction.
Section VIII — Battery Replacement
See Figure 21 and do the following to replace the Battery Pack:
• Power down the unit.
• Loosen the Battery Door screw, on back of pump.
• Open Battery Door and disconnect wire harness. A small straight-bladed screw driver may be needed to disconnect
the connector tab.
• Slide Battery Pack out of compartment.
• Slide a new Battery Pack into compartment with wire harness extending out.
• Align connector with red wire at upper right.
• Push connector in until properly latched.
• Tuck wires into slot between connector and battery to insure wires are not pinched. Re-tighten Battery Door screw.
Note:
Replacement Battery Packs are available from Tyco Healthcare/Kendall Co. (See Section XIII - Service Part
Numbers).
WARNING:
USE OF A BATTERY PACK NOT SUPPLIED BY TYCO HEALTHCARE/KENDALL CAN BE DANGEROUS ANDVOIDS
ALL WARRANTIES AND PERFORMANCE SPECIFICATIONS.
The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is required
to fully recharge the Battery Pack.
Figure 21. Installation of Battery Pack.