25
Enteral Feeding Pump
English
English
Lock Screen Feature
The KANGAROO ePump enteral feeding pump can be set to
RUN
and then locked out from unauthorized use. While the
LOCK SETTINGS
feature prevents only changes to the input parameters, the
LOCK SCREEN
feature will lock out any
button presses, including
RUN
or
HOLD
, except power off.
Use
LOCK SCREEN
when the pump is running and there is a need to prevent tampering such as putting the pump into an
unwanted
HOLD
. The
RUNNING
screen will show a padlock to indicate that no changes can be made without unlocking
RUNNING
RUNNING
the screen.
To lock the screen, use the following procedure while the pump is in
RUNNING
mode:
RUNNING
RUNNING
• Press and hold the second
4
button from the top for at least 2 seconds, and then release it.
• Press and hold the third
4
button from the top for at least 2 seconds, and then release it.
• Press and hold the second
4
button again for at least 2 seconds.
To unlock the screen, use the same procedure above.
Section VI – Re-Certification of Performance
System Performance Re-Certification
A series of self-diagnostic checks can be run to re-certify pump performance. It is recommended that this re-certification
procedure be run at least once per year, or any time the pump is suspected of having improper performance. See Section
XIII-Service Part Numbers for ordering new Re-Certification Pump Sets.
Figure 19. Program Locked screen,
which indicates the pump is in LOCK
SETTINGS mode.
Figure 20. Padlock
indicates pump is in
LOCK SCREEN mode.
© 2005 Tyco Healthcare Group LP