General Startup
Placement/Mounting
There are two recommended placement methods for the KANGAROO ePump enteral feeding pump.
• Attached to a vertical IV pole via the Pole Clamp peripheral device, included with the pump.
• Placed feet down on any horizontal surface.
A/C Power Operation
Plug the KANGAROO ePump enteral feeding pump into an A/C power outlet for normal operation. When the pump is
not plugged in, or if A/C power is interrupted, a built-in rechargeable battery will run the pump.
Battery Power Operation
If A/C power is not available to the pump, or if A/C power is interrupted, the pump will automatically operate on backup
battery power. The built-in battery recharges any time the pump is plugged into an outlet.
A fully-charged, new battery will supply 15 hours of backup power at 125 ml/hr before requiring recharging. Contact
Customer Service, Section XI, to replace the Battery Pack if battery life degrades.
Power On/Off
To power up the pump, press the
POWER
button on the lower right of the front panel. To power down the pump,
press and hold the
POWER
button until the pump display turns off.
Language Selection, First Power Up
Immediately after powering up the pump for the first time, the language option screen will appear. Select the language
using the (
È
Immediately after powering up the pump for the first time, the language option screen will appear. Select the language
Immediately after powering up the pump for the first time, the language option screen will appear. Select the language
) or (
Í
Immediately after powering up the pump for the first time, the language option screen will appear. Select the language
Immediately after powering up the pump for the first time, the language option screen will appear. Select the language
) buttons. When the language is selected, push the fifth button (
Á
) to enter the language
selection. After the language is selected the first time the pump is powered up, the language selection screen will not
appear during future power up operations. Should the language need to be changed after the initial power up, it can be
changed through the “
MORE OPTIONS
” menu (see Figure 16).
Keep or Clear Prior Pump Settings
Immediately after powering up the pump, if the prior feed or flush settings were not cleared and the settings are not
locked in the BioTech Screen, the pump will give the following two options:
KEEP SETTINGS
- Select this option to start with the same settings that were most recently programmed into the
pump. Any of the previously-programmed settings can be modified, if needed. The feed volume totals, “
mL FED
”
(Continuous Feed Mode) or “
BOLUSES FED
” (Intermittent Feed Mode) and “
mL FLUSH
” are also maintained.
CLEAR SETTINGS
- Select this option to reset all input feeding settings and the output feed totals to zero. It will then
be necessary to program all settings before running the pump.
Note:
The output feed totals (“
mL FED
,” “
,,
BOLUSES FED
,” “
mL FLUSH
”) can be cleared at any time, and without
clearing the input settings, by doing the following:
• Start the pump
RUNNING
• Press
4
HOLD
• Press
4
CLEAR VOLUME
The status line of the pump will then show
LOAD A SET
(Figure 5), or, if the Pump Set is already loaded, it will show
A SET
A SET
SET LOADED
(Figure 7).
SET LOADED
SET LOADED
To power up the pump, press the
To power up the pump, press the
press and hold the
press and hold the