Continuous Mode (Adjust Feed)
Continuous Mode Feeding
For
CONTINUOUS
feeding mode, the
ADJUST FEED
option is available in the opening menus (Figure 7A for Feed-
Only sets or Figure 7B for Feed & Flush sets). Select this option to program the feeding parameters from the
ADJUST
FEED
screen, Figure 10A.
Press
4
FEED RATE
in Figure 10A to define the rate of delivery. Press the
4
buttons in Figure 10B to program the rate
from 1 to 400 in increments of 1 ml/hr. Press
4
ENTER
(top button) to exit the menu.
Press
4
FEED VTBD
in Figure 10A to define the Volume To Be Delivered. Press the
4
buttons in Figure 10C to
program the volume from 1 to 3000 ml in increments of 1 ml. Press
4
ENTER
(top button) to exit the menu. If the
FEED VTBD
is set to zero, the pump will run until the supply is exhausted.
NOTE:
When a
FEED VTBD
is set, the remaining volume to be delivered (remaining VTBD) will be displayed next to
the v icon on the lower right-hand portion of the screen (Figure 10A). If the
FEED VTBD
is reprogrammed during
a feeding, from
HOLDING
mode, the pump will restart the feeding and deliver the entire volume of the new VTBD
HOLDING
HOLDING
setting.
Figure 9A.
MORE OPTIONS
menu.
MORE OPTIONS
MORE OPTIONS
Figure 9B.
SELECT MODE
menu for
SELECT MODE
SELECT MODE
continuous or intermittent mode
selection.
Selecting Continuous or Intermittent Feed Mode
The KANGAROO ePump enteral feeding pump can be programmed to feed continuously or intermittently.
The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has
been delivered or until the supply has been exhausted. The Intermittent feeding mode delivers boluses of enteral nutrition
at programmed time intervals. The bolus volume and feed rate are also programmed.
To select continuous feed mode or intermittent feed mode, press
4
MORE
on the opening menu, shown in Figures 5 or 7,
to access the
MORE OPTIONS
menu, Figure 9A.
MORE OPTIONS
MORE OPTIONS
The
MORE
option is available either before a Pump Set has been loaded (Figure 5) or after the Pump Set is loaded (Figure
7). See subsection “More Options” for information on the other options in the
MORE OPTIONS
menu.
MORE OPTIONS
MORE OPTIONS
Select
4
CONTINUOUS/INTERMITTENT
under the
MORE OPTIONS
menu, Figure 9A. Then press
4
CONTINUOUS MODE
or
4
INTERMITTENT MODE
from the
SELECT MODE
menu, Figure 9B. Press
SELECT MODE
SELECT MODE
4
DONE
to
exit.