18
KANGAROO
R
R
e
Pump
™
English
English
Intermittent Mode (Adjust Bolus)
Intermittent Mode Feeding
For
INTERMITTENT
feeding mode, the
ADJUST BOLUS
option is available in the opening menus (Figure 7C for
a Feed-Only set or Figure 7D for a Feed & Flush set). Select this option to program the feeding parameters from the
ADJUST BOLUS
screen, Figure 12A.
ADJUST BOLUS
ADJUST BOLUS
Press
4
BOLUS RATE
in Figure 12A to define the rate of delivery.
Press the
4
buttons in Figure 12B to program the delivery rate from 1 to 400 in increments of 1 ml/hr.
Press
4
ENTER
(top button) to exit the menu.
Press
4
NUMBER OF BOLUSES
in Figure 12A to define the total number of boluses to be delivered.
Press the
4
buttons in Figure 12C to program the number of boluses, up to a maximum of 99.
Press
4
ENTER
(top button) to exit the menu. The number of boluses can be set as zero, in which case the pump will
run until the supply is exhausted.
Press
4
VOLUME PER BOLUS
in Figure 12A to define the Volume to be delivered in each bolus.
Press the
4
buttons in Figure 12D to program the volume per bolus from 1 to 1000 ml in increments of 1 ml.
Press
4
ENTER
(top button) to exit the menu.
Press
4
BOLUS INTERVAL
in Figure 12A to define the time interval between starts of bolus deliveries.
Press the
4
buttons in Figure 12E to program the time interval between starts of bolus deliveries from 1 to 24 hours in
increments of 1 hour. Press
4
ENTER
(top button) to exit the menu.
The pump will automatically limit bolus settings which exceed pump capabilities to deliver. For example, a bolus
delivery volume of 200 mL cannot be made once per hour if a rate of 150 mL/hr is already programmed. In this case,
the maximum volume allowed would be 150 mL.
NOTE:
If the
NUMBER OF BOLUSES
is reprogrammed during a feeding (from HOLD mode), the pump will restart
the bolus feeding and will deliver all the boluses as defined by the new setting.
© 2005 Tyco Healthcare Group LP