24
KANGAROO
R
R
e
Pump
™
English
English
BioTech Screen Data
SN
is the serial number of the pump.
SW
shows the software version number in this pump.
FLASH
shows the version number of the flash data installed in this pump.
CERT INTV
indicates the number of times the pump has been powered on since the last re-certification. See Section
CERT INTV
CERT INTV
VI for details of re-certifying the pump.
Lock Settings Feature (Program Locked)
To “lock” the settings against unauthorized changes, press
4
LOCK SETTINGS
to get the check mark, as shown inFigure
18. The settings can only be changed again by pressing
4
LOCK SETTINGS
to remove the check mark. If the settings
are locked and a change is attempted, the screen will indicate
PROGRAM LOCKED
, and will not allow access to the
parameters, Figure 19.
The pump can be
PRIMED
or set to
RUN
,
HOLD
, etc. when settings are locked.
Press
4
DONE
to exit the BioTech menu. To re-enter this menu screen, power down the pump and then power it up
again and hold the top
4
button, as described above.
BioTech Menu
The BioTech screen will show technical information about the pump, and has a lock feature to lock the Pump Settings
against unauthorized changes.
Accessing the BioTech Screen
To get the BioTech screen, press the power button to start the pump, and press and hold the top
4
button while the
startup sequence (Kangaroo icon “hops” across the screen) executes. You must be pressing the top
4
button at the end of
the sequence in order to access the BioTech menu. Figure 18 shows the BioTech screen.
Figure 18. BioTech screen, with
Pump Settings “locked” against
unauthorized changes.
© 2005 Tyco Healthcare Group LP