35
Enteral Feeding Pump
English
English
Flushing Solution Dose Range
10-500 ml in 1 ml increments
Flushing Solution Interval Range
1-24 hr in 1 hr increments
Accuracy
±10% or 0.5 ml/hr, whichever is larger, with bag at 46 cm (18”) above pump, at room temperature 22° C ± 2° C (72°
F ± 3° F), using a new pump set with less than the recommended 24 hours of maximum usage.
Occlusion Pressure
15 psi (103 kPa) Maximum
Dimensions
Height: 16.8 cm (6.6”) Width: 16.3 cm (6.4”) Depth: 11.7 cm (4.6”)
Weight
1.1 kg (2.4 lbs), 1.2 kg (2.7 lbs) with pole clamp
Material
Housing: ABS/Polycarbonate blend
Pump Set Door: Polyester/ Polycarbonate blend
Power
Use AC adapter for wall outlet usage. Pump operates on 9V DC, 1.5 Amps.
Battery
New, fully charged NiMH Battery Pack delivers ≥ 15 hrs at 100 ml/hr feed rate. Approximately 15 minutes prior
to complete battery discharge, a low battery alarm will occur (see subsection “Battery Low” in Section IX). When
complete discharge occurs, the pump will automatically turn itself off.
The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is
required to fully recharge the battery pack.
Alarms
• System Error
• Flow Error
• Pump Set Dislodged Error
• Hold Error
• Rotor Error
• Feed Error
• Flush Error
• Battery Low
Alarm Volume
Minimum of 70 dBA at 1 meter in maximum volume orientation
© 2005 Tyco Healthcare Group LP