15
Enteral Feeding Pump
English
English
Auto Priming
The
AUTO PRIME
option will be available if the pump senses that the currently installed Pump Set was not previously
primed, and that there is no fluid in the line near the rotor. If auto-priming is desired, press and release
4
AUTO PRIME
to automatically prime the line(s). For Feed & Flush Pump Sets, the auto prime feature will automatically prime both
lines, starting with the flush line.
Press
4
STOP
to cancel
AUTO PRIME
.
When auto-priming is completed, the status line will show
AUTO PRIME COMPLETE
and will no longer show the
AUTO PRIME COMPLETE
AUTO PRIME COMPLETE
AUTO PRIME
option. Be sure that the line(s) are fully primed. If not, use the hold-to-prime option(s) to complete the
priming, as described below.
Hold-to-Prime
The hold-to-prime options allow for precise interactive control of the priming process.
Feed-Only Pump Sets
For Feed-Only Pump Sets, the screen will appear as in Figure 8A. Press and hold
4
HOLD TO PRIME FEED
until the
feed line has been primed down to the stepped end connector at the end of the Pump Set.
Feed & Flush Pump Sets
For Feed & Flush Pump Sets, the display will appear as in Figure 8B. First press and hold
4
HOLD TO PRIME FLUSH
to
prime the flushing line, and then press and hold
4
HOLD TO PRIME FEED
to prime the feed line down to the stepped
end connector at the end of the Pump Set.
If a Feed & Flush Pump Set is used, remember to first press
4
HOLD TO PRIME FLUSH
until the flushing fluid has
reached the valve, and then press
4
HOLD TO PRIME FEED
until the feeding fluid has reached past the valve and
down to the stepped end connector. If the feeding line is primed first, subsequent priming of the flushing line will force
the air between the flush solution bag and valve into the main line, which already has been primed with enteral formula.
Re-priming after Bag Empty
A Pump Set bag that has been emptied will trigger the
FEED ERROR
screen. In this condition the Pump Set bag can be
refilled to continue the feeding, but only after the Pump Set has been re-primed.
To re-prime the Pump Set, do the following:
• Disconnect the feeding line from the patient
• Refill the bag
• Press
4
CONTINUE
to begin the pump
RUNNING
• Press
4
HOLD
• Press
4
ADJUST SETTINGS
• Press
4
PRIME PUMP
• Press
4
HOLD TO PRIME
to prime the line(s)
Pump Sets should not be reused after 24 hours of initial usage.
© 2005 Tyco Healthcare Group LP