Clear Volume Fed
Press
4
CLEAR VOLUME
to re-zero the output feed totals “
mL FED
,” “
BOLUSES FED
,” and “
mL FLUSH
.”
Adjust Settings
Press
4
ADJUST SETTINGS
to return to one of the opening menus shown in Figure 7, thus allowing any setting to be
modified. If the Pump Set has been removed after going into
HOLD
, the opening menu will show
LOAD A SET
, as seen
LOAD A SET
LOAD A SET
in Figure 5.
Resume After Specified Time Interval
Press
4
RESUME IN __ MIN
to immediately input a time in minutes in which to resume operation. Each press of
4
RESUME IN __ MIN
will increment the resume time by 5-minute increments, up to 30 minutes. After “30” is
reached, each additional button press will increment in larger time steps, up to a maximum of 240 minutes. Another
button push after 240 will cycle the “resume in” time back to zero. No other input is needed to start the countdown, as
the time immediately counts down after the incremental value is reached.
Run Mode
Press
4
RUN
to immediately restart pump operation. This option will not be available if the Pump Set is removed during
HOLD
.
View History
Press
4
HISTORY
to access the
HISTORY
HISTORY
HISTORY
HIST
HIST
screen. Use this screen to review feeding and flushing history. See subsection
ORY
ORY
“More Options” below for further information.
Flush Now
During the
RUNNING
mode, the pump can be diverted to immediately perform a flush (flush on demand) by
RUNNING
RUNNING
pressing
4
FLUSH NOW
in the
RUNNING
menu. See Figure 14.
RUNNING
RUNNING
When
4
FLUSH NOW
is pressed, the screen
will prompt for the volume to be immediately flushed, and will show the previously programmed flush volume as
default. The screen will appear as the standard Flush input screen as shown in Figure 11B (or 13B). Setting the
FLUSH
NOW
flushing volume is temporary and will not change the main/periodic flushing volume previously programmed.
If
4
FLUSH NOW
is pressed by accident, the screen will return to the
RUNNING
screen after 10 seconds, if there is no
RUNNING
RUNNING
input for the flush volume.