4 Firmware description
86
Idle state
The unit is waiting for an examination to be started by the
driver. The service state can be entered during this state
The unit will enter the error state if a system error occurs.
Select (program select) state
The unit initializes the selected program and enters the
ready state when system is ready to take an exposure. If
the examination is ended, the unit will return to the idle
state. The unit will enter the error state if a system error
occurs.
Ready state
The unit is waiting for the exposure button to be pressed.
Once it notices that the button has been pressed the
exposure state is entered.
If the exposure program is changed when the unit is in the
ready state, the unit will revert to the select state. The unit
will enter the error state if a system error occurs.
Exposure state
The unit carries out the selected exposure program and,
when completed, proceeds to the transfer (image transfer)
state.
The exposure button is monitored during the exposure.
Rotation and sensor movements are also monitored. If a
fatal error occurs and no image has been captured the unit
will enter the error state. If an image has been captured the
unit will enter the transfer state and transfer the image to
PC and then enter the error state.
Transfer (image transfer) state
The unit waits for the image transfer to be completed. Any
calibration actions are handled if the exposure program
was a calibration program.
The exposure counter is updated.
After the driver has received the image it releases it.
It makes sure that exposure button is released and after it
has been the unit returns to the select state.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...