7 Error codes and troubleshooting
150
SOLUTION E
Use service program “readadc” to determine
whether the temperature is measured correctly.
- ADC value 111 corresponds to the maxi-
mum allowed temperature (75 deg Celsi-
us). Lower ADC values indicate a
incorrectly calibrated measurement. Ad-
just trimmer R66 in the EG200 XRAY IF
board until approximately 200 is read as
ADC value at room temperature.
- ADC value 255 indicates the temperature
sensor (connected to EG200/JG217) be-
ing loose or broken. Check connection and
replace the sensor if broken. The resist-
ance in the sensor should be roughly 4.7
kOhm.
Generator temperature
sensor
is broken
CAUSE A
The generator temperature sensor is broken or
loose.
SOLUTION A
Check that the generator temperature sensor is
properly connected. The sensor is connected to
JG217 in the EG200 X-RAY IF board and should
have a resistance of about 4.7 kOhm at room tem-
perature.
Service period has expired and unit need maintenance
CAUSE A
The service reminder timer has expired (by default
one year after last service).
SOLUTION A
Carry out annual service as lined out in section 9
Maintenance Procedure. Mark service as done us-
ing the “indicator” s2terminal service command.
End user quality assurance is needed to retain image quality
CAUSE A
The quality assurance timer has expired (by default
180 days after last 3D QA).
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...