7 Error codes and troubleshooting
167
7.6.7.2
Exposure sequence starts but no X-rays are
generated
•
Check the condition of the power supply by follow-
ing the trouble shooting guide 7.6.3.8 for R3220
and R3300.
•
Set the unit to imaging mode (CT, Pan or CEPH),
disable movements using command
nomove
(s2terminal) and check status of the PREH_ENA
LED (D6).
If LED D6 does not come on, check the
condition of the cable between J3202
and JG206.
If the cable is OK, replace
EG200.
If not, replace cable.
If LED D6 comes on, the exposure switch
is OK.
Follow high voltage trouble
shooting.
If LED D6 does not come on, measure the volt-
age, JG206-6, during preheat.
Replace EG200 if the voltage is about
5V.
Measure voltage from J3202-6 during
preheat.
Replace cable between JG206
J3202 if there is voltage on
J3202.
Replace R3220 if there is no volt-
age on the J3202.
If LED D6 comes on, check the EXP_ENA LED
D7.
If LED D7 comes on, follow the trouble
shooting guide for X-ray generation.
If LED D7 does not come on, measure
the voltage during exposure from con-
nector J3303 pin 5.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...