7 Error codes and troubleshooting
153
SOLUTION A
Check if the diagnostic value of the resulting image
is sufficient. Take another image if needed.
Exposure button is pressed
CAUSE A
The exposure button is pressed after imaging has
ended or during startup.
SOLUTION A
Release the exposure button.
CAUSE B
Exposure button in stuck down, shorted or broken.
SOLUTION B
Check the exposure switch with a multimeter, re-
place if faulty.
CAUSE C
Faulty cable between EA300 Extension board
(JA301) and EA200 Connector board (JA204).
SOLUTION C
Check the cabling. Replace if needed.
CAUSE D
Faulty EA200 Connector board, EA300 extension
board or R3400 IO board.
SOLUTION D
Try to locate the failing board using a multimeter.
Replace faulty board(s).
Nasion support is in wrong position
CAUSE A
The cephalostat nasion support (ear holders) is not
rotated correctly for the selected imaging program.
SOLUTION A
Rotate the nasion support into correct position.
CAUSE B
The nasion support position detection is not detect-
ed correctly by the cephalostat CPU.
SOLUTION B
Check that the position detection micro switch is
activated / deactivated properly when the nasion
support is rotated. Check the cable between the mi-
cro switch and EG200/JG204. Check the serial
data line between EG200/JG205 and R3220/
J3213. Replace micro switch, cables or EG200 if
needed.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...