7 Error codes and troubleshooting
160
7.6.3.2
X-, Y- or rotator not working
•
Check +5V and +24V voltage indicators on the
stepper modules.
If not, replace EA100.
If after replacing EA100
the LEDs still do not come
on, replace the low power
supply.
If LEDs are on after un-plugging
some of the connectors, then fol-
low the power distribution chain in
question by cutting it step-by-step
until the LED in question goes
out. Replace the next component
(cable or board).
If each LED on EA100 is on then check the LEDs
at the end of each power chain (see block dia-
gram)
If some LED are not on, localize the
failure as described above in
“introduction”
If the LEDs are on replace R3220 main
CPU board
If both indicators are on, follow the motion control
trouble shooting guide.
If +5V indicator LED is not on but R3300 is active,
replace the cable between stepper module
(J5101) and R3300.
If LED still does not come on, replace
R5100
If +24V indicator LED is not on, check that +24V
indicator LED on the EA100 is on.
If not, see trouble shooting instructions in
chapter 7.6.3.1.
If indicator is on, m24V voltage
from J3309.
Replace cable between J3309
and JA106 if there is no voltage.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...