56
3.5.8
R5300 Panel raw power supply
DESCRIPTION:
There is one board in each base unit but two if the CEPH
unit exists.
The first one gen8V low voltage for the analog side
of the 3D and panorama sensor power chains.
The second one gen8V low voltage for the analog
side of the CEPH sensor power chains.
NOTE!
The +8V is distributed to the EG400 power supply,
which re-regulates it down to +6V for panorama / CEPH
sensors and down to +5V for the 3D sensor.
LOCATION:
First one is located in the vertical carriage for the 3D and
panorama sensors.
Second one is located in the CEPH unit for the CEPH sen-
sor.
INPUTS:
12Vac
OUTPUTS:
+8Vdc
(analog side)
FUSES:
F1
T2A/250V 6 x 32
INDICATOR LEDS:
H1
= 12Vac (input)
H2
= +8V (output)
TEST PINS:
None
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...