7 Error codes and troubleshooting
124
7.4.4
Error 9 R3300 IO1 board is not
functioning
CAUSE A
Communication between the main CPU and R3300
is not working properly.
SOLUTION A
• The Ethernet cable between the R3300 and the
internal Ethernet switch may be faulty. Try with a
spare cable and replace the cable if needed.
• If also error 12 is shown, check the cable between
the main CPU and the internal Ethernet switch. Try
with a spare cable and replace the cable if needed.
• Check the internal Ethernet switch and replace if
needed:
- Appropriate leds are lit: each port having a
cable connected should have the link led
on.
- The specified voltage level is present in
power connector of the switch.
CAUSE B
The R3300 IO board software is incompatible with
the main CPU software. State commanded for
beeper is not matching the feedback value.
SOLUTION B
Update the software(s). Verify that the versions list-
ed by the s2terminal command “login” match ver-
sions.txt in the software update package.
CAUSE C
The R3300 IO board is faulty and therefore not re-
sponding.
SOLUTION C
Replace the board.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...