7 Error codes and troubleshooting
138
7.4.26
Error 41 Movement of T (cephalostat
sensor) motor failed
CAUSE A
There is a problem with the T position detection.
SOLUTION A
Use service program “cephstatus” to check the en-
coder status. Move the sensor back and forth and
verify that the encoder value changes. To resolve
an encoder or opto switch error, check the cable
between the opto board and the cephalostat CPU.
Replace parts if needed.
CAUSE B
The T movement is not working correctly.
SOLUTION B
The movement not working correctly may be due to
a fault in the
• motor,
• motor cable,
• R5100 Stepper driver module,
• cable between the stepper driver module and the
cephalostat CPU board.
Use service command “drivemotor T manual” to
test whether the motor is working properly. Check
for cables or other parts hindering the movement.
Inspect cables and connectors, and replace parts
one by one if needed.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...