7 Error codes and troubleshooting
143
7.4.34
Error 49 Reading panoramic sensor
data failed
CAUSE A
The sensor is not properly connected.
SOLUTION A
Lift the lever at the side of the panoramic sensor,
and firmly push it back.
CAUSE B
The connection between the sensor and the sensor
holder is bad.
SOLUTION B
Power off the unit and clean the connectors in the
sensor and sensor holder with flux remover. Make
sure the connectors are dry before powering on
again.
CAUSE C
Voltage level out of range or voltage missing.
SOLUTION C
Enable power for the panoramic sensor using ser-
vice command “panelpower panoramic 1”. Measure
the digital voltage level from pin 1 and 2 on the sen-
sor’s power connector. The allowed range is
4.9…6V. If the voltage is missing, locate the point of
failure with the help of the schematic in 5.4.2 Pano-
ramic imaging chain. Replace faulty parts.
CAUSE D
Faulty / loose data cables or connector boards.
SOLUTION D
Verify that all cables related to the panoramic imag-
ing chain are intact and properly connected.
CAUSE E
Faulty panoramic sensor.
SOLUTION E
Replace the sensor and redo all panoramic calibra-
tions.
1. Analog GND
2. 6V
3. Digital GND
4. D6V
1.
2.
3.
4.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...