7 Error codes and troubleshooting
141
7.4.31
Error 46 Emergency stop button is
pressed
CAUSE A
The emergency stop button is activated.
SOLUTION A
Release emergency stop button and dismiss the
error message. If the unit is equipped with a ceph-
alostat, emergency may also be activated from the
cephalostat.
CAUSE B
The emergency stop switch is not working properly.
The switch shall open the circuit when pressed and
make a short circuit otherwise.
SOLUTION B
Check the switch operation with a multimeter and
replace if faulty. If the unit is equipped with a ceph-
alostat, check also the emergency stop button lo-
cated in the bottom of the cephalostat.
CAUSE C
The loopback connector shorting the cephalostat
emergency stop for non-ceph units is missing (shall
be connected to JE006 in the top of the pillar).
SOLUTION C
Attach the loopback connector, or if it’s present,
check that the emergency stop signal is passing
through the connector properly.
CAUSE D
Emergency stop switch cabling is faulty.
SOLUTION D
Short the pins in JA303 and JA304 in the EA300
Extension board. If the error is no longer present,
the cables are faulty and need to be repaired/re-
placed. If shorting the pins in the J304 removes the
error, the problem is related to the cephalostat (ca-
bles, emergency switch, connectors). Check and
repair/replace cables if needed.
CAUSE E
EA300 extension board or R3400 IO board is
faulty.
SOLUTION E
If the error is still present with the pins in JA303 and
JA304 shorted, the error is in the EA300 or R3400
board. Replace both boards.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...