7 Error codes and troubleshooting
131
7.4.19
Error 34 Movement of D (sensor)
motor failed
CAUSE A
The sensor is not rotating correctly.
SOLUTION A
The movement not working correctly may be due to
a fault in the
• motor,
• motor cable,
• R5100 Stepper driver module or the
• cable between the stepper driver module and the
CPU board.
In rare cases the Main CPU may also be faulty.
Use service command “drivemotor d manual” to
test whether the motor is working properly. If the
collimator is moving, the problem is probably in po-
sition detection ( see CAUSE B). Otherwise one of
the components listed above is faulty. Continue
with inspecting cables and connectors. Replace
parts one by one if needed.
CAUSE B
Position detection showing incorrect values.
SOLUTION B
Use service command “drivemotor d manual” and
the chin support up/down buttons to drive past the
positions where the opto/micro switches are acti-
vated/deactivated. Check for messages in the
s2terminal indicating the switches changing state.
If no such messages are seen, there is a problem
with the EG100 opto board, the sensor position mi-
croswitch MSW2 or the cable between the switches
and the EG200 XRAY IF board. If also motors C
and N are failing, the serial IO data cable between
EG200/J205 and R3220/J3213 may be faulty.
Check the parts listed and replace if needed.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...