7 Error codes and troubleshooting
156
7.6 Hardware trouble shooting
7.6.1
Internal and external LANs
7.6.1.1
No Ethernet connection between the unit
and the workstation
•
See if the problem can be resolved by restarting
the X-ray unit and the workstation.
•
Check for problems in the workstation.
•
Check for problems in network cables.
Try to set up a connection between the X-ray unit
and the workstation using the s2terminal.
If a connection is established, check the
installation and configuration of the image
acquisition driver software.
Try to set up a connection between the X-ray unit
and another PC (for example service
laptop) using the s2terminal. If a connection can
be established the problem is with the
workstation.
Check that the “link” light, normally located next
to the workstation’s network connector, is on
when the cable is attached. If it is not on, the
network card may be faulty. Check if the light is
on when the service laptop is connected, this
confirms whether the network card is faulty or
not.
Replace the network cable between the X-ray
unit and the workstation with an Ethernet cable
(CAT5e or CAT6) known to be working.
If a connection is established, replace the original
cable.
Connect the known to be working Ethernet cable
between the workstation and connector JB004 in
the upper part of the column. If a connection is
established, replace the energy chain cables lo-
cated behind the column.
Connect the known to be working Ethernet cable
between the workstation and connector J3220
on the R3220 Main CPU board. If a connection is
established, replace the whole main cabling in
the rotating unit.
Summary of Contents for ORTHOPANTOMOGRAPH OP 3D Pro
Page 2: ......
Page 4: ......
Page 18: ...1 Introduction 10...
Page 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Page 31: ...2 Unit description 23...
Page 32: ...2 Unit description 24...
Page 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Page 54: ...46...
Page 87: ...79...
Page 88: ...80...
Page 96: ...4 Firmware description 88...
Page 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Page 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Page 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Page 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Page 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Page 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Page 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Page 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Page 122: ...6 The s2terminal and service functions 114...
Page 192: ...8 Maintenance procedure 184...