Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.1 Control unit service program
ON/OFF Optimized Feedback
Meaning: Turn the optimised feedback on/off
The command "ON/OFF Optimized Feedback" activate/deactivate the optimised
feedback function. The status is displayed in the status bar (1).
Note
This menu item is only for clinical tests and should generally be set to OFF.
Detect/Feedback/US en/disable
Meaning: Activate or deactivate detect and feedback mode. the Detect and Feed‐
back mode can be activated or deactivated by entering two codes.
These codes depend on the serial number of the KEY Laser 1243.
The activation codes for all KEY Lasers from SN 101 to SN 220 are listed in a table
in the service update "Exchanging the control computer starting software version
2.15".
Starting with SN 221, the codes are on a sticker near the energy measuring head.
Press F stands for the feedback code, and press D stands for the detect code.
If the KEY Laser is purchased in the USA, the sticker is removed during production.
If the codes have been lost, they can be obtained from the laser hotline if you provide
the serial number.
67/176