Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.1 Control unit service program
▶ ENERGY power meter serial no.: = Enter the serial number of the energy power
meter and choose ENTER to confirm.
▶ Measuring head serial no.: = Enter the serial number of the measuring head and
choose ENTER to confirm.
▶ Sensitivity: = enter the sensitivity of the energy meter, and confirm with ENTER.
▶ Confirmed factors A, B and C (0, 1 and 0) if not otherwise indicated.
The selection menu "Commissioning adjustments" appears.
Resonator adjustment
This function is used to adjust the resonator
Note
Follow all assembly instructions.
Note
The laser power supply unit, the cooling system, the compressor and the flash lamp
switch on automatically.
▶ Choose OK to confirm the Resonator adjustment command.
The laser pulse generator is on standby together with the drive for 80 mJ/6 Hz.
▶ Use the key "ON/OFF Laser" to start or stop the laser pulse generator.
▶ Choose EXIT to return to the Commissioning adjustments menu.
Deflection mirror adjustment
This function is used to adjust the deflection mirror
Note
Follow the description of startup on the assembly instructions.
58/176