Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.2 Optical system
① grip sleeve Mat. no. 1.001.0845
② coupling (not a replacement part)
③ lens holder (not a replacement part)
④Threaded ring key Mat. no. 0.411.0042
⑤ spacing ring (delivered with the adjusted lens)
⑥Adjusted lens Mat. no. 1.001.0848
⑦Threaded ring Mat. no. 0.574.2052
Note
The spacing ring ⑤ and the lens ⑥ are adjusted to each other in the plant and may
only be exchanged together.
Transmission check
If you want to check the energy at the output of laser handpiece E 2062, the trans‐
mission factor must be taken into account.
The transmission factors for the various fibre plugs and frequencies can be found
in the "Instructions for use of KEY Laser handpieces"
A distance of 30 mm between the energy meter and fiber plug is measured.
The output is calculated as follows:
Energy set at the display X transmission factor = energy at the output
Example (Fibre plug 30/38, 160mJ/10Hz):
160mJ x 0.33 = 50mJ
144/176