Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.1 Control unit service program
Shutter/Absorber test
This is the shutter and absorber test
Choose OK to confirm the Shutter/Absorber test command.
The shutter and absorber will open and close alternately.
The switching time and feedback signals will be checked.
The test indicates the set target position ① and the actual position detected ②
1
2
Definitions
Absorber reset position ➨ swivelled in!
Shutter reset position ➨ swivelled in!
The term “swivelled in“ indicates that the laser beam is interrupted.
Swivel out absorber Swivel out shutter
The term “swivelled out“ indicates that the laser beam is released.
Error detection
1. If the shutter and/or absorber remains defective in the closed position, the mes‐
sage ➨ Position error ➨ Timeout error for the shutter and/or absorber aperture is
displayed under ②.
▶ Press the swing arm switch.
2. If the shutter and/or absorber remains defective in the open position, the message
Absorber/Shutter reset Position ➨ swivelled out ➨ Position error message is dis‐
played.