Technician's instructions KEY Laser III 1243
3 Maintenance work on KEY Laser 1243
|
3.11 KEY Laser complete
3.11 KEY Laser complete
Component
Jobs to be done
Exchange interval
Parts to replace as needed
Check the power cable for
damage.
Power cable
Mat. no. 1.000.7704
Check the foot button ca‐
ble for damage.
Hose Mat. no. 0.226.9164
Starter hose
Mat. no. 1.000.4482
Calibration
Check or perform calibra‐
tion steps according to in‐
stallation instructions.
Identify flash lamp wear as
follows: If the control va‐
lues need to be increased
by more than 20% in the
energy adjustment, ex‐
change the flash lamp.
Compare the control ta‐
bles of the first and present
energy adjustments. A
prerequisite is that no re‐
placement part influencing
the output energy was ex‐
changed between the first
and present energy adjust‐
ment.
Flash lamp
Mat. no. 1.000.7711
Feedback system
Check and adjust the feed‐
back system according to
the service update.
Safety checks.
The safety checks and do‐
cumentation must follow
assembly instructions.