Technician's instructions KEY Laser III 1243
3 Maintenance work on KEY Laser 1243
|
3.5 Laser contraangle handpiece 2060
3.5 Laser contraangle handpiece 2060
Component
Work to be performed
Replacement interval
Parts to be replaced if ne‐
cessary
Spray
Unscrew spray nozzle and
clean fixing bore.
Replace spray O-
ring
Assembled nozzle
Mat. no. 1.000.4849➊
O-ring
Mat. no. 0.200.6013➊
Replace O-ring on stopper O-ring
Mat. no. 0.200. 6120 ➊
Check spray pattern of in‐
strument and make sure it
is airtight.
ATTENTION: Also in‐
wards towards the optical
system.
Optical system
Replace O-ring on exit
window.
O-ring
Mat. no. 0.200.6120 ➊
Check exit window.
Assembled window
Mat. no. 1.000.4847.
Perform a visual inpsecti‐
on of the deflection mirror
and, if necessary, clean
according to instructions
Coated mirror
Mat. no. 1.000.4852
Check spray and transmis‐
sion.
For details, see the Setup
instructions and User in‐
structions for handpieces.