Instructions for use ESTETICA E70 Vision / E80 Vision
3 Product description | 3.10 Technical data
42 / 178
Operating environment
WARNING
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
▶
It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
Floor quality
The quality of the flooring must meet
the load bearing ability for buildings
DIN 1055 page 3 and have a pres-
sure resistance in accordance with
DIN 18560 T 1.
Ambient temperature
+10 to +40
o
C
Optimum ambient temperature
15°C to 35°C
Relative Humidity
30% to 75% non-condensing
Air pressure
700 hPa - 1,060 hPa
Max. altitude for operation
up to 3000 m
Maximum loads
Maximum patient weight load
180 kg
Tray holder of the dentist element -
loadable up to
2 kg
Assistant element standard tray -
loadable up to
1 kg
Dentist element - loadable up to
2 kg
Service table 1568 without locking
system
2 kg
Service table 1568 with locking sys-
tem
5 kg
Transportation and storage conditions
Ambient temperature
-20 to +55
o
C / -4 to +131°F
Relative Humidity
5% to 95% non-condensing
Air pressure
700 to 1060 hPa