Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.18 Use USB interface
4.17.4 Replacing the high-pressure bulb of the
COMFORTbase
CAUTION
Danger of burns from hot high-pressure lamp.
▶
Switch main device switch off.
▶
Let the COMFORTbase cool down after long use.
Requirement
The COMFORTdrive is pulled off from the COMFORTbase.
▶
Push the accompanying lamp changer on the high-pressure lamp and pull
out the lamp axially.
▶
Insert the new lamp into the lamp changer, and introduce it into the hole in
the face of the supply hose. Carefully slide the lamp into the mount.
▶
Carefully press out the lamp by activating the ejector.
4.17.5 Replacing the KaVo MULTI LED lamp
See also:
2
4.13.4.2.2 Replacing the KaVo MULTI LED lamp, Page 112
4.18 Use USB interface
CAUTION
Electrical power.
Electrical shock.
▶
Set up the external PC outside of the patient environment keeping a min-
imal distance of 1.5 m.
▶
Connect the PC and equipment connected to the PC in accordance with IEC
60601-1 / 60950.
141 / 178