Instructions for use ESTETICA E70 Vision / E80 Vision
2 Safety | 2.3 Safety instructions
18 / 178
CAUTION
Third-party device connection assembly kit (optional): Risk of reverse
contamination by stagnant water.
Infections.
If a water consumer is connected to the third-party device connection as-
sembly kit, always perform the following actions on the device:
▶
Before starting, rinse all the water drain lines with no handpieces attached
(if applicable).
▶
Before initial start-up and after downtimes (weekends, holidays, vacations,
etc.), rinse or purge the air and water lines with air.
▶
Mind the H
2
O
2
resistance of the water consumer since the water is doped
with OXYGENAL 6 cleaner (concentration up to 0.02 %).
CAUTION
Health damage due to germ formation.
Infection hazard.
▶
Before starting, rinse all the water drain lines without instruments.
▶
Before start-up and after the device has not been used for a while (week-
ends, holidays, vacations, etc.), rinse or purge with air the air and water
lines.
▶
Carry out an intensive germ reduction.
▶
Actuate the tumbler filler repeatedly.
CAUTION
Risk of injury and material damage from incorrect use of the charger
for the wireless foot control.
Personal injuries, damage to the wireless foot control or the charger.
▶
Do not use the treatment unit during the charging process!
▶
Do not use the wireless foot control charger supplied to charge non-re-
chargeable batteries.
▶
Charge the wireless foot control with the charger supplied only.
CAUTION
Long stay in the patient chair.
Decubitus formation.
▶
Take precautions against the formation of decubitus in long treatments.
CAUTION
Danger of injury from tipping the treatment unit.
Injury to the patient and user.
▶
Do not support yourself on the swinging arm.
▶
Do not sit on the head or foot end of the patient chair when it is in a hori-
zontal position.
CAUTION
Danger of injury from overload or dynamic load.
The patient chair might collapse.
▶
Do not subject the patient chair to a load exceeding its limit (180 kg).
▶
Do not subject the patient chair to dynamic loads.