Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.15 Using the SL600 surgical motor (optional accessory)
4.15.4 Calling-up the surgery mode
Requirement
The surgical motor is connected.
No handpiece has been taken off the holders.
▶
Unfold the "Treatment mode" selection list and select "Surgery".
4.15.5 Mount or pull off the handpiece or contra-angle
handpiece
Note
Follow the instructions for use, service instructions and installation instruc-
tions in the instrument packaging.
NOTICE
Damage from changing the handpiece and angle piece during opera-
tion.
Wear to the catch on the straight and contra-angle handpiece and motor.
Unbalanced motor axis.
▶
Change the handpiece and angle piece only when the motor not running.
Attaching the straight or contra-angle handpiece
All straight and contra-angle handpieces with INTRAmatic connection ISO 3964
can be attached.
▶
Place the KaVo handpiece on the motor, lightly press it against the motor
while turning it slightly in the direction of the arrow until the guide stud can
be heard to lock into place.
▶
Pull on the KaVo handpiece to make sure that it is securely attached to the
motor.