Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.13 Using instruments
112 / 178
▶
Pull off the grip sleeve
④
together with the cannula from the valve body
①
.
Replacing the high-pressure lamp
▶
Push the holder
③
towards the front and draw and remove the defective
high pressure lamp
②
from its holder.
▶
Install a new high pressure lamp (
Mat. no. 1.002.2928
).
Replacing the KaVo MULTI LED lamp
Note
The KaVo MULTI LED bulb is a semiconductor element and may only be oper-
ated with direct current. To ensure proper function, the poles need to be in-
serted correctly.
▶
Push the holder
③
forward and pull the defective KaVo MULTI LED lamp
②
out of the socket.
▶
Insert new Kavo MULTI LED lamp (
Mat. no. 1.007.5372
).
The following may happen after you turn on the KaVo MULTI LED lamp:
▪ Case 1: KaVo MULTI LED lamp is on.
▪ Case 2: KaVo MULTI LED lamp is faint.
- Increase the cold light intensity on the unit until the desired light intensity
is reached.
▪ Case 3: KaVo MULTI LED lamp is red or off.
- Take KaVo MULTI LED lamp out of its socket as described above and re-in-
sert it after rotating it 180° about its axis.
4.13.5 Using the PiezoLED
CAUTION
Handpiece inserts can be damaged from long-term use, or when
dropped or bent.
They cannot be guaranteed to function properly.
Injury from insert breakage.
▶
Check the handpiece inserts before each use.