Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.12 Using the foot control
4.12.12 Use physiological saline solution (optional
accessory)
Requirement
Treatment centre is turned on. The handpiece is connected to the pump via
the pressure line.
▶
Remove the handpiece.
▶
Press the cross-switch of the foot control for 4 seconds until you hear the
signal.
▶
After activation, select the "NaCl" cooling on the dentist control unit.
4.12.13 Charge the wireless foot control
The wireless foot control is operated by means of an installed rechargeable bat-
tery.
CAUTION
Risk of injury and material damage from incorrect use of the charger
for the wireless foot control.
Personal injuries, damage to the wireless foot control or the charger.
▶
Do not use the treatment unit during the charging process!
▶
Do not use the wireless foot control charger supplied to charge non-re-
chargeable batteries.
▶
Charge the wireless foot control with the charger supplied only.
Note
Charge the wireless foot control with the charger supplied by KaVo only.
Note
The wireless foot control charger may only be used indoors and must be pro-
tected from moisture.
▶
Connect the charger to the wireless foot control.
The charger display communicates the following:
Display
Meaning
shines green
The unit is ready
shines yellow
The battery is being charged
Shines weak green
Battery is charged
does not shine
The battery is dead, or there is a short-circuit
The battery voltage is above the tolerance range
Reversed poles
The transition from charging to full is indicated by the fluttering of the display.