Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.16 Use pump for physiological saline solution (optional accessory)
Icon
Description
Expiry date
Batch number
Sterilisation method
Single use
4.16.2 Connecting the coolant
Note
All liquid-conducting parts are not sterile. They must be sterilised before the
first treatment. All parts conducting liquids must be kept sterile.
See also:
2
Servicing instructions
Connect the coolant with the standard instrument hose
▶
Attach the pressure line
①
to the motor hose with the enclosed hose clips
②
.
Note
The distance from the motor to the first hose clip must be approx. 80 mm.
Note
The enclosed hose clips fit the motor hose of the surgical motor S600 LED.
Use the hose clips
Mat. no. 0.211.7492
for the motor hose of the INTRA
LUX KL 703 LED.
135 / 178