Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.16 Use pump for physiological saline solution (optional accessory)
Note
One-touch calibration should be carried out only with KaVo surgical hand-
pieces with a transmission ratio of 16:1 or 20:1.
The one-touch calibration cannot be run with third-party handpieces or hand-
pieces with different transmission ratios.
CAUTION
The motor starts at full speed.
Risk of injury.
▶
Hold the motor firmly or put it in a safe holder during the calibration.
▶
Press the foot control and hold it down until the display shows a message
indicating that "Calibration was successful".
▶
If you release the foot control before the display shows that the calibration
was successful, press the foot control again until the display shows a mes-
sage indicating that the calibration was successful.
▶
Tap the "Cancel" key to terminate the calibration and to return to the selec-
tion of device settings.
If an unsuitable or defective handpiece was used in the calibration, the calibra-
tion is discontinued and the error message, "Measurement failed - Non-per-
missible current", is shown.
▶
Tap the "Cancel" key to terminate the failed calibration.
4.15.11 Exit surgery mode
▶
Tap the "Cancel" key to exit from "Surgery" as the type of treatment.
4.16 Use pump for physiological saline solution
(optional accessory)
Note
A sterile saline solution can be used using the physiological saline solution
pump instead of the treatment water from the unit. The application of the
pump is designed for instruments with a suitable interface for the coolant.
133 / 178