5 Configuration
92
5.17
Batches
Batch logging is configured here (available as of system version 6).
The batch data is displayed together with the channels of the assigned groups and the event messages
in the batch visualizations (evaluation with JUMO smartWARE Evaluation).
The maximum number of batches that can be recorded depends on the device (variTRON 300: 10 batch-
es; variTRON 500: 20 batches). However, the number of licensed batches is decisive for the evaluation
of the recorded batches (fee-based licenses for JUMO smartWARE Evaluation).
NOTE!
If no group has been assigned to the batch (or none of the assigned groups is active), only event mes-
sages are recorded.
NOTE!
With the variTRON 300, the user must ensure that the maximum number of 10 batches is not exceeded.
If the maximum number of batches is exceeded during configuration, this will result in the batch in ques-
tion not being recorded. In case of configuration with the Web Cockpit, a fault message is immediately
entered into the event list.
NOTE!
The configuration must not be changed during an ongoing batch reporting process. Otherwise, there is
a risk that the batch documentation will become unusable.
General information
Batch lines
For each batch, up to 10 lines can be defined for the batch report.
Parameter
Selection/data input
Description
Batch active
Off (FALSE)
Batch reporting is not active.
On (TRUE)
Batch reporting is active and can be started.
Batch application
(batch name)
Enter text (or use existing text) Assignment or designation of the batch
Control signal
Select signal from selector
Signal that is used to start (high active) and stop
batch reporting.
Response to
power interrup-
tion (behavior af-
ter power
interruption)
Continuation
Batch reporting is continued.
Abort
Batch reporting is stopped.
The batch is completed with the last recorded data
(before the power failure).
Group allocation
Select group from selector
Group that is assigned to the batch.
All groups can be assigned to one batch.
Parameter
Selection/data input
Description
ID
Enter text (or use existing text) Designation (ID) of the relevant line of the batch re-
port (e.g. for use in the PLC application)
The identifier must be unique within a batch.
Text left column
Enter text (or use existing text) Text in the left column of the batch report.
The left column contains the designation for the
content of the right column.
Summary of Contents for 705003
Page 2: ......
Page 6: ...Contents...
Page 20: ...1 Introduction 20...
Page 25: ...25 2 Mounting 2 3 Dimensions 101 110 5 45 94 5 101 5 17 5...
Page 26: ...2 Mounting 26...
Page 68: ...4 Operation 68 The files are available for download in the web browser click on the file...
Page 94: ...5 Configuration 94...
Page 98: ...6 Startup 98...
Page 109: ...109 7 Annex 7 5 China RoHS...
Page 110: ...7 Annex 110...
Page 111: ......